In many technical, industrial, or scientific fields, testing is a central component of quality assurance. However, a measurement result alone does not indicate whether a product or component actually meets the specified requirements. This assessment is carried out as part of what is known as conformity assessment. In this article, we explain in easy-to-understand terms how a laboratory arrives at a conformity statement in accordance with ISO/IEC 17025:2018 and DAkkS requirements. CleanControlling describes the procedure for assessing the conformity of test results in document FB 35-74, which forms the basis for accreditation.
Essentially, conformity assessment answers a simple question:
"Does the measured result meet the required specification – yes or no?"
Specifications may include, for example:
To do this, the laboratory compares the measured value with the specified limit or target value. The result of the evaluation is always clear:
✔ compliant (fulfilled) or
✘ non-compliant (not fulfilled)).
In some cases, customers only want the measured values without any statement regarding conformity. This is useful, for example, in pure status quo analyses or development projects.
In such cases, the following applies:
The test report clearly states that no conformity statement has been made.
A decision rule specifies how the laboratory deals with measurement uncertainty in conformity assessment.
Why this is important:
Every measurement involves a certain degree of uncertainty. A decision rule defines a range within which the true value is highly likely to lie.
Example:
A measurement result is 98, the limit value is 100, and the measurement uncertainty is ± 5
→ The true value could therefore be between 93 and 103.
The decision rule answers the question:
"Is measurement uncertainty taken into account in conformity assessment or not?"
Three possible approaches are described:
If there is a binding decision rule (e.g., in a standard), the laboratory applies only this rule.
Customers can also define their own specifications in writing, e.g.:
The laboratory then explicitly documents this in the test report.
If no specifications exist, the laboratory uses simple acceptance:
Only the measured value counts—the measurement uncertainty is NOT taken into account for the evaluation.
This means:
The measurement uncertainty is nevertheless determined and stored internally.
The document FB 35-74 "Conformity of Test Results" from CleanControlling shows typical situations when dealing with limit values in the "simple acceptance" procedure. Figure 1 shows the seven cases (A to G) at a glance.
It is important to note that in the standard procedure, only the measured value counts, not the uncertainty. Therefore, cases C and D are still considered compliant, even though the uncertainty reaches the limit value.
The documentation required by ISO/IEC 17025:2018 ensures that:
The transparent presentation makes it clear:
Conformity assessments are always systematic and rule-based — not subjective.
Conformity assessment is an essential step in correctly classifying measurement results and reliably meeting quality requirements.
This procedure ensures that:
This forms the basis for transparent, high-quality auditing services.
Newsletter registration